| Coatron M1 | Coagulation Analyzer | ||
|---|---|---|---|
| General | |||
| Dimensions | 245 x 130 x 65 mm | ||
| Weight | 0.55 kg | ||
| Power Supply (external): | |||
| Input | 100-240 V~, 50/60Hz | ||
| Power | 12W | ||
| Approvals | CE, EN60950, UL | ||
| Incubation Block | |||
| Temperature | 37,0°C ± 0,2°C | ||
| Samples | 6 cuvettes | ||
| Reagent | 2 vial Ø11,0 | ||
| Different adapters available | |||
| Measurement | |||
| Optic | 1 channel with AutoSense | ||
| Methods | Turbidity, Kinetic, Endpoint | ||
| Algorithm | based on optical density | ||
| Processing | 14Bit ADC + Digital Filter | ||
| Autosense | LED intensity controlling | ||
| Starting | manual or optical | ||
| Performance | |||
| Precision Throughput | CV < 5% PT ~25/h APTT ~10/h FIB ~17/h | ||
| Test | |||
| Global | PT, aPTT, Thrombintime, Fibrinogen (Clauss & PT derived), Fibrinogen | ||
| Factors | Factor | ||
| Thrombophilia | Protein C, Antithrombin, Ecarin Chromogenic Assay for Thrombin, Ecarin Chromogenic Assay for Hirudin, quantitative D-Dimer (immunturbidimetric) | ||
| Calculated units | Seconds, INR, %, U (mg/dL or ng/mL) | ||