Release 1.04.52 Edition-2 12.05.2025 Software Changes: CR1: (only for OEM version=1) New test methods for reagent C1010-020 TEChrom anti-Xa , chromogenic assay) -AF10: Screening of anti-Xa activity -APIX: Apixaban level -RIVA: Rivaroxaban level -EDOX: Edoxaban level -HEP: Heparin (default for LMWH) Major Fixes Compatibility problems after update of older firmware 1.02.xx solved, like -RESET DEVICE -CHANGE SIN -Internal barcode-scanner does not switch Compatibility: The new version can be installed on all instruments and versions without any limitations. At first reboot after update, all test protocols and calibration data be reset to default. --------------------------------------------------------------------------------- Release 1.04.52 Edition-1 11.07.2024 Software Changes: CR1: Support of next cuvette voucher generation, which will come to market 2024-07. Older vouchers can still be used. CR2: Support of 'Code25' encryption, which will be introduced together with next cuvette voucher generation. CR3: Improved onboard result history , which allows statistical analysis and reporting of patient's and quality control results (QC1 & QC2). CR4: (only for OEM version=1) New test methods: -PC Protein-C level (C1100-007 TEChrom PC, chromogenic-assay) -FPS free Protein-S level (A600-002 Protein S, immuno-assays) -VWF von Willebrand Factor (immuno-assays) CR5: A service tag is display on status screen, which is required for support requests CR6: Support of "Hi-Dose" option for immunoassays such as DD,FPS,VWF. It allows reflex testing for very high samples with enhanced linearity. Note: This change is only available for TECO Coatron X Series, but not for variants. Compatibility: The new version can be installed on all instruments and versions without any limitations. At first reboot after update, all test protocols and calibration data be reset to default. --------------------------------------------------------------------------------- Release 1.03.51 08.08.2022 Global changes none OEM changes 1F - method HEP is activated 2B - method DD is activated Minor Fixes x System reported red status "CLOCK" and displayed "SYSTEM NOT READY". Any reset of system date to current date was rejected. Fixed now. x Utra turbid suspensions like Kaolin suspension can cause optical problems. The automatic light intensity adjustment is now optimized for this use-case. x The default expiry date of all tests was 0/00. It took around 20min to scroll from default 0/00 to Aug-2022. Fixed now. --------------------------------------------------------------------------------- Release 1.03.50 30.03.2020 OEM=all CR1: Easy Input of calibration curve. First point (eg. 100%, 12.5s) must be entered manually, but further points are assisted now. CR2: Hidden test protocol parameter "Limit" can be set down to 0% ! In this case the instrument break at an early state of clot-reaction and report the time at start of clot. This feature is helpful for assays like Protein-S where the start and endpoint of clot reaction differs more than 60s. CR3: Hidden test protocol parameter "Start" can be set now down to 2s. This is helpful for extreme fast starting reactions like with FIB. CR4: New languages onboard English,Deutsch,Espanol,Italian,Francais,Romana,Polski,Dansk, Nederlands,Srpski,Eesti keel,Suomi,Hrvatski,Latviesu,Lietuviu, Portugues,Svenska,Slovensky,Slovenscina,Cestina,Magyar, Minor Fixes x Serial communication optimizied for with TECMONI V2.97. x Sometimes valid barcodes were rejected. x Results out of range ( e.g. >250%) were reported only in seconds x A change of hidden test protocol is now valid for both LOT1 and LOT2 of each test. Note: After update following data will be cleared and reset to factory default. Test Calibration Data: Yes, all tests Test Protocol Data: Yes, all tests Result Database: No System settings: No --------------------------------------------------------------------------------- Release 1.03.49 24.01.2020 OEM=all CR1: Statistics for results Open result database. There is now a "QC" button, which filters current PID &Test and print mean result as well as C.V. value. CR2: DOAC test scroll button Run any DOAC test ( API,RIVA,EDOX,LMH or AF10) and swipe current result through all anti-Xa agents like Apixaban,Rivaroxaban,Edoxaban. CR3: Test default protocol index introduced for each OEM 1.03.49 (OEM-1D) Major=1 Minor=3 Build=49 OEM=1 Protocols="D" CR4: Green color of optic button after test-start indicates idle time. Here it is allowed to touch,.move or mix cuvette. After idle time the color switch to red . CR5: System accepts now alphanumeric patient-ID at barcode input. CR6: System sends now mean result to TECAM (LIS) in mode "Double" CR7: Incubation alarm continues counting to negative time Minor Fixes x Temperature changed upto 0.5 C during measurement. x Calibration curves were printed without decimalpoint like 6g/l instead of 6.00g/l --------------------------------------------------------------------------------- Release 1.02.48 09.07.2019 OEM=all Minor fixes: Test was automatically reset to default calibration, if system date exceeded the reagent expiry date. Fixed now. No calibration data will be changed. --------------------------------------------------------------------------------- Release 1.02.47 03.06.2019 OEM=all CR1: Results range (Min - Max) can be limited to calibration range. Then results are not extrapolated, but reported as Max. CR2: No-clots are reported as >maxtime instead of "+++" CR3: Polish dialogue language is enabled. CR4: Support of new TECAM Smart V7.06.0 It allows to visualize and manage test calibration data with TECAM. Major fixes: System got nearly frozen, if connected to TECAM and if reagent lotnumber is set to not-zero. --------------------------------------------------------------------------------- Release 1.02.46 6.2.2019 CR1: Bugfix versus false test units printed on screen (eg. "%" instead og "mg/dl") The problem was since Build-45. CR2: Digital wavefilter included for chromogenic or immunturbidimetric tests By was filter the C.V. of low DD was improved from 15% down to 9%, By was filter the C.V. of highDD was improved from 8% down to 2.5%, CR3: PT-% is now display with one decimal digit CR4: System supports now Code20-tickets, which is required for use of vouchers in year 2020 --------------------------------------------------------------------------------- Release 1.01.45 19.11.2018 CR1: Test PC+PS+LA-c+LA-S added CR2: HiSense Button only visible for clotting tests CR3: DD Default protocol changed to START=60 Stop=180s CR4: Result expression for unit "R" and "NR" is now activated Minor Fixes Note:After update the system will reset to default and clear -all calibration data -all stored results --------------------------------------------------------------------------------- Release 1.01.44 07.05.2018 CR1: New language "Romana" CR2: Bugfix of chromogenic methods. With Build-43 always 0.034E was measured --------------------------------------------------------------------------------- Release 1.01.43 21.03.2018 CR1: Heating during standby removed CR2: Input of cuvette ticket with alphanumeric keyboard CR3: Mirror print function to TECAM SMART --------------------------------------------------------------------------------- Release 1.01.42 19.03.2018 First Release