Release 1.04.52  Edition-2  12.05.2025

Software Changes:

CR1:   (only for OEM version=1) 
New test methods for reagent C1010-020 TEChrom anti-Xa , chromogenic assay)
 -AF10:  Screening of anti-Xa activity
 -APIX:  Apixaban level
 -RIVA:  Rivaroxaban level
 -EDOX:  Edoxaban level
 -HEP:   Heparin (default for LMWH)

Major Fixes
Compatibility problems after update of older firmware 1.02.xx solved, like
  -RESET DEVICE 
  -CHANGE SIN
  -Internal barcode-scanner does not switch

Compatibility:
The new version can be installed on all instruments and versions without any limitations.
At first reboot after update, all test protocols and calibration data be reset to default.

---------------------------------------------------------------------------------
Release 1.04.52  Edition-1  11.07.2024

Software Changes:

CR1:   Support of next cuvette voucher generation, which will come to market 2024-07.
Older vouchers can still be used.
		 
CR2:   Support of 'Code25' encryption, which will be introduced together with 
next cuvette voucher generation.

CR3:   Improved onboard result history , which allows statistical analysis and reporting of patient's 
and quality control results (QC1 & QC2). 	

CR4:  (only for OEM version=1) 
New test methods: 
-PC  Protein-C level (C1100-007 TEChrom PC, chromogenic-assay)
-FPS free Protein-S level (A600-002 Protein S, immuno-assays)
-VWF von Willebrand Factor (immuno-assays)
	
CR5:   A service tag is display on status screen, which is required for support requests

CR6:   Support of "Hi-Dose" option for immunoassays such as DD,FPS,VWF.
It allows reflex testing for very high samples with enhanced linearity.
Note: This change is only available for TECO Coatron X Series, but not for variants.

Compatibility:
The new version can be installed on all instruments and versions without any limitations.
At first reboot after update, all test protocols and calibration data be reset to default.

---------------------------------------------------------------------------------
Release 1.03.51  08.08.2022

Global changes
       none

OEM changes
       1F  - method HEP is activated
       2B  - method DD  is activated

Minor Fixes
       x System reported red status "CLOCK" and displayed "SYSTEM NOT READY".
         Any reset of system date to current date was rejected. Fixed now.   
       
       x Utra turbid suspensions like Kaolin suspension can cause optical problems.
         The automatic light intensity adjustment is now optimized for this use-case.

       x The default expiry date of all tests was 0/00. It took around 20min to scroll
	 from default 0/00 to Aug-2022. Fixed now.
		
---------------------------------------------------------------------------------
Release 1.03.50  30.03.2020 OEM=all

CR1:   Easy Input of calibration curve. First point (eg. 100%, 12.5s) must be entered manually,
       but further points are assisted now.

CR2:   Hidden test protocol parameter "Limit" can be set down to 0% !
       In this case the instrument break at an early state of clot-reaction and report the time at start of clot.
       This feature is helpful for assays like Protein-S where the start and endpoint of clot reaction differs more than 60s.
 
CR3:   Hidden test protocol parameter "Start" can be set now down to 2s.
       This is helpful for extreme fast starting reactions like with FIB.

CR4:   New languages onboard
       English,Deutsch,Espanol,Italian,Francais,Romana,Polski,Dansk,
       Nederlands,Srpski,Eesti keel,Suomi,Hrvatski,Latviesu,Lietuviu,
       Portugues,Svenska,Slovensky,Slovenscina,Cestina,Magyar,
	
Minor Fixes
       x Serial communication optimizied for with TECMONI V2.97.
       x Sometimes valid barcodes were rejected.  
       x Results out of range  ( e.g. >250%)  were reported only in seconds
       x A change of hidden test protocol is now valid for both LOT1 and LOT2 of each test.

Note:  After update following data will be cleared and reset to factory default.
	Test Calibration Data:	Yes, all tests
	Test Protocol Data:	Yes, all tests
	Result Database:	No
	System settings:	No

---------------------------------------------------------------------------------
Release 1.03.49  24.01.2020 OEM=all

CR1:   Statistics for results
       Open result database. There is now a "QC" button, which filters current PID &Test and
       print mean result as well as C.V. value.

CR2:   DOAC test scroll button
       Run any DOAC test ( API,RIVA,EDOX,LMH or AF10) and swipe current result through all anti-Xa agents like Apixaban,Rivaroxaban,Edoxaban.
	
CR3:   Test default protocol index introduced for each OEM 1.03.49 (OEM-1D)
       Major=1  Minor=3 Build=49 OEM=1 Protocols="D"

CR4:   Green color of optic button after test-start indicates idle time. Here it is allowed to touch,.move or mix cuvette.
       After idle time the color switch to red .

CR5:   System accepts now alphanumeric patient-ID at barcode input.

CR6:   System sends now mean result to TECAM (LIS) in mode "Double"

CR7:   Incubation alarm continues counting to negative time

Minor Fixes
       x Temperature changed upto 0.5 C during measurement. 
       x Calibration curves were printed without decimalpoint like 6g/l instead of 6.00g/l

---------------------------------------------------------------------------------
Release 1.02.48  09.07.2019 OEM=all
      
Minor fixes:
      Test was automatically reset to default calibration, if system date exceeded
      the reagent expiry date. Fixed now. No calibration data will be changed.

---------------------------------------------------------------------------------
Release 1.02.47  03.06.2019 OEM=all


CR1:  Results range (Min - Max) can be limited to calibration range. 
      Then results are not extrapolated, but reported as <Min or >Max.

CR2:  No-clots are reported as >maxtime instead of "+++"

CR3:  Polish dialogue language is enabled.

CR4:  Support of new TECAM Smart V7.06.0 	
      It allows to visualize and manage test calibration data with TECAM.
      
Major fixes:
      System got nearly frozen, if connected to TECAM and if reagent lotnumber 
      is set to not-zero.

---------------------------------------------------------------------------------
Release 1.02.46   6.2.2019

CR1: Bugfix versus false test units printed on screen (eg. "%" instead og "mg/dl")
     The problem was since Build-45.	


CR2: Digital wavefilter included for chromogenic or immunturbidimetric tests
     By was filter the C.V. of low DD was improved from 15% down to 9%,
     By was filter the C.V. of highDD was improved from 8% down to 2.5%,


CR3: PT-% is now display with one decimal digit

CR4: System supports now Code20-tickets, which is required for use of vouchers 
     in year 2020

---------------------------------------------------------------------------------
Release 1.01.45   19.11.2018

CR1: Test PC+PS+LA-c+LA-S added
CR2: HiSense Button only visible for clotting tests
CR3: DD Default protocol changed to START=60 Stop=180s
CR4: Result expression for unit "R" and "NR" is now activated

Minor Fixes

Note:After update the system will reset to default and clear 
-all calibration data
-all stored results

---------------------------------------------------------------------------------
Release 1.01.44   07.05.2018

CR1: New language "Romana"
CR2: Bugfix of chromogenic methods. With Build-43 always 0.034E was measured

---------------------------------------------------------------------------------
Release 1.01.43   21.03.2018

CR1: Heating during standby removed
CR2: Input of cuvette ticket with alphanumeric keyboard
CR3: Mirror print function to TECAM SMART

---------------------------------------------------------------------------------
Release 1.01.42   19.03.2018

First Release